B.R. Nahata College of Pharmacy Mhow- Neemuch Road, 458001 Mandsaur, Madhya Pradesh, India.
*E-mail: anurekha@brncop.com
https://doi.org/10.53879/id.50.05.p0048
ABSTRACT
A validated high performance liquid chromatographic method was developed for the determination of chromium picolinate in pharmaceutical dosage forms. The analysis was performed at room temperature using a reversed-phase ODS, 5µm (250×4.6) mm column. The mobile phase consisted of acetonitrile: buffer (60:40 V/V) at a flow rate of 0.5 mL/min. The PDA-detector was set at 264 nm. The developed method showed a good linear relationship in the concentration range from 1.5 – 12.5 µg/mL with a correlation coefficient from 0.999. The limit of detection and limit of quantification were 0.0540513 and 0.1637919 µg/mL respectively.